Calibration
15
3.3 Calibration Overview
Remove the unit from the case by loosening all four captive fasteners on the front
panel and lifting the panel out of the case, then do the following:
1. Center the fine tuning (page 16).
2. Turn off the scaling (page 16).
3. Adjust for no load (page 18).
4. Adjust for a known or calculated load (page 19–standard calibration;
page 21–shunt calibration).
5. Adjust the scaling (page 19–standard calibration; page 21–shunt
calibration).