![Honeywell NK Скачать руководство пользователя страница 16](http://html.mh-extra.com/html/honeywell/nk/nk_user-manual_151866016.webp)
Setup and Use
12
2.8 Fine Tune
Fine tune the Model NK (the controls for both
zero
and
span
are on the front
panel):
1.
Fine zero
—With no load on the transducer, use a small screwdriver to turn
the
zero
potentiometer until the display reads
0000
.
2.
Fine span
—With the known load on the transducer, use a small
screwdriver to adjust the
span
potentiometer until the display matches
your known load.
OR
With no load on the transducer, press and hold the
shunt cal
button. Use a
small screwdriver to adjust the
span
potentiometer until the display
matches the value calculated on page 11.
Note: Turn the potentiometer clockwise to increase the displayed value. Turn it
counterclockwise to decrease the displayed value.