Calibration
18
3.6 Adjust for No Load
1. Remove any load from the transducer.
2.
Coarse zero
—This control zeroes the display when the
fine zero
is out of
range or the transducer has a preload, offset, or tare.
Use a small screwdriver to turn the
coarse zero
potentiometer until the
display reads approximately
0000
(Figure 4 on page 17).
3.
Fine zero
—Use a small screwdriver to turn the
fine zero
potentiometer on
the front panel until the display reads exactly
0000
.
Note: Turn the potentiometer clockwise to increase the displayed value. Turn it
counterclockwise to decrease the displayed value.