10.7. Coated aluminium parts
Rilsan®: PA 11 = Blue color
The coating should be subject to regular and careful inspection. The continued used of the apparatus should
not be permitted if inspection reveals that the protective material has become damaged to the extend that the
underlying protected metal is visible, until such damage has been satisfactorily repaired.
10.8. Winding action becoming stiff
If after repeated use the winding action is becoming slightly stiff apply the following simple process:
−
engage the tape cleaner (position "UP"), with the sensor retained in the storage tube,
−
slacken the tape a few turns, typically 10,
−
gently shake the instrument to free up the tape within the tape housing,
−
wind the tape again and disengage the tape cleaner (position "DOWN").
10.9. Electrical checking of the tape assembly
⇒
Test for grounding
−
Remove the battery holder as described in
section 10.3.
⇒
Measure the resistance between the ground (-)
terminal (as shown on Figure 11-1) of the
electronic circuit and the tube of the sensing
probe; the resistance should be less than 10
Ω
. If it is higher, the steel tape might be broken
or the connection between the sensing probe
circuit and the sensing probe tube might be
interrupted.
⇒
Test for short-circuit
−
Disconnect the tape at both ends: display unit
side and sensing probe side (see sections
−
Measure the resistance between each
conductor red-white, red-black, white-black.
This resistance should be infinite as an open
circuit. If not, the tape might be defective.
⇒
Test for open-circuit (continuity)
−
Disconnect the tape at the sensing probe side
−
Measure the resistance of each conductor of
the tape (between red and red, white and
white, etc.).
−
The resistance should be less than 15
Ω
. If not,
the tape might be broken. To replace the tape
see section 10.4.
Figure 11-1
Содержание HERMetic UTImeter Gtex Chem IIC
Страница 5: ...50489 GTEX Chem IIC 1805 5 UTImeter Gtex Chem IIC Figure 2 2 ...
Страница 15: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 15 Figure 7 2 ...
Страница 55: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 55 Figure 13 1 general assembly list of the main spare parts ...
Страница 56: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 56 Figure 13 2 display unit assembly TS 10227 details ...
Страница 59: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 59 Figure 13 5 storage tube SS1 Q1 TS 10186 details ...
Страница 60: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 60 Figure 13 6 storage tube SS1 Q2 TS 10187 details ...
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......