GENESIS DM CLINICIAN GUIDE
Installing the Genesis DM Monitor
P.4510EN.01
12/5/2007
37
Honeywell HomMed and Fairbanks scales are interchangeable provided you have the appropriate
scale connector cable. If you exchange scales:
•
Read the label on the scale connector cord to verify that you have the correct scale cable for
the scale you intend to attach to the monitor.
•
Power-down the monitor
before you exchange scales. Attach and position the new scale
before
you turn the monitor back on.
Attaching the blood pressure cuff
Push the metal connection on the blood pressure tubing firmly into the NIBP port. It should click when
it locks into place.
Attaching the modem
CAUTION
REQUIRES ANALOG PHONE LINE
The modem requires an analog phone line. A digital phone line may destroy the
modem.
DO NOT use a digital phone line. Verify presence of analog phone line prior to
installation.
The modem connection is on the back of the Genesis DM monitor.
To connect the modem to the phone outlet, follow the steps below.
1.
Unplug the patient’s phone line from the wall telephone jack.
2.
Plug the provided duplex jack into the wall jack.
3.
Plug the patient’s telephone into one side of the duplex jack.
4.
Plug the Genesis DM phone cord into the other side of the duplex jack.
5.
Plug the remaining end of the Genesis DM phone cord into the
MODEM
plug located on the back
of the monitor.
6.
Alternately, you can connect the GPRS modem to connect wirelessly.