GENESIS DM CLINICIAN GUIDE
Programming the Genesis DM Monitor
P.4510EN.01
12/5/2007
33
Option Menu
The
Options Menu
allows you to assign communication ports to compatible peripheral medical devices.
You must assign a com port to each device or the monitor
will not prompt
for the device and it
will
not collect data
from that device, even if the device is connected to the com port.
After assigning a com port, you can then program how often the Genesis DM collects test results from
the device. The default is daily. However, you can program any combination of days. You must
have at least one day selected or the monitor will not prompt for the device.
Service Menu
This menu allows Honeywell HomMed technicians to access the Genesis DM monitor to obtain specific
diagnostic data and perform calibration tasks. If you are asked to perform any tasks that require you to
enter this menu, Customer Service will walk you through the steps.
Service Call
Choose this item to initiate a connection between the monitor and the Honeywell HomMed service
department computer.
Service Call Phone # (C)
This item contains the number the monitor uses to connect to the Honeywell HomMed service
department computer.
Calibration Verify (C)
This item is used to perform a calibration test on the NIBP measurement.
WARNING
RISK OF INJURY
Verifying calibration while the cuff is attached to a patient could cause bruising or other injury.
DO NOT verify NIBP calibration while the cuff is attached to a patient.
Report Menu (C)
Choose the Report Menu (C) to access the following information:
Serial Number (C)
This item reports the serial number assigned to the monitor.