Operation
5.3 Unit Reset
Overview
Unit Reset initializes all of the DirectLine
®
Sensor’s calibration and configuration
data to factory default values. The Air calibration value is also reset to the factory
default.
Procedure
From the Online DO display, press and hold the
and
buttons
simultaneously until “
rSEt
” appears on the display (
minimum of 10 seconds
).
“
rSEt
” will remain on the display for about 8 seconds followed by the firmware
version, the DO Measurement units (ppm or ppb) and the online DO
Concentration value. The unit then returns to the Online DO display.
Table 5-4 Factory Default Values
Data
Default Values
Air/Sample calibration
Factory default
Atmospheric Pressure
760 mm Hg
Salinity
0 ppt
Probe Bias Voltage
0.55 Volts
Noise Suppression Frequency Selection
60 Hz
Output Configuration – 0 % Range Value
4.00mA
Output Configuration – 100 % Range Value
20.00mA
Output Configuration – 0 % Calibration
0.00 ppm / 0.0 ppb
Output Configuration – 100 % Calibration
20.00 ppm/200 ppb
26
DirectLine® DL424/425 Sensor Module User Manual
9/11