Configuration
9/11 DirectLine®
DL424/425 Sensor Module User Manual
17
Step
Operation
Press
Display
5
Enter
Output
Configuration
MODE
Momentarily
OutC Enter Output Calibration
100% Range Value
Selection
Momentarily
rnGH
(
for 1 second
) then,
(V
alue of current 100 % Range Value
Selection
)
Edit 100% Range
Value Selection
MODE
Hold
Flashing Display – You are now in EDIT
mode
(V
alue of current 100 % selection
)
Select desired
100% DO Range
Momentarily
Selected 100% DO Value
Range:
DL424 = 0 to 20 ppm
DL425 = 0 to 200 ppb
Save the New 100%
Range Value
MODE
Momentarily
(
New Value
)
6
0 % Range Value
Selection
Momentarily
rnGL
(
for 1 second
) then,
(
value of current 0% Range Value
Selection
)
Edit 0 % Range
Value Selection
MODE
Hold
Flashing Display
– You are now in EDIT
mode
(
value of current 0 % selection
)
Select 0 % DO
Value
Momentarily
Selected 0 % DO Value
Range:
DL424 = 0 to 20 ppm
DL425 = 0 to 200 ppb
Save the New 0 %
Range Value
MODE
Momentarily
(
New Value
)
7
100 % Calibration
Momentarily
AdJH
Adjust 100 %
Calibration
MODE
Hold
AdJH (flashes) – You are now in EDIT
mode
Range: 19.60 to 20.40 mA typically
(
default 20.00 mA
)
Momentarily
+AdJH
(increments value)
AdJH
(decrements value)
Save 100 %
Calibration
MODE
Momentarily
AdJH