Configuration
3.2 Configuration Set Up Procedure
ATTENTION:
In Table 3-2, under the
Press
column:
Hold
means to hold the button down until the display changes.
Momentarily
means to press and release the indicated button.
From the Online DO display, follow this procedure.
ATTENTION:
If no key is pressed for 60 seconds, the display will abort the entry mode and
default to Online Display.
Table 3-2 Configuration Set Up Procedure
Step
Operation
Press
Display
1
Enter
Atmospheric
Pressure
Compensation
MODE
Hold
AtPr
(
for 1 second
), then
Value of Atmospheric Pressure Value in
mm Hg
Edit Atmospheric
Pressure Value
MODE
Hold
Flashing Display – You are now in EDIT
mode
(Value of Current Atmospheric Pressure
Compensation )
Select a new value
or
Momentarily
Enter a value from 500 to 800 mm Hg (760
mm HG default)
Save the value
MODE
Momentarily
Saves the new Atmospheric Pressure
Compensation value for use at the next Air
Calibration.
2
Enter
Salinity
Compensation
(ppm units only)
MODE
Momentarily
SALn
(
for 1 second
), then
Value of current Salinity Compensation in
Parts per Thousand
Edit Salinity
Compensation
MODE
Hold
Flashing Display – You are now in EDIT
mode
(Value of current Salinity Compensation)
Select a new value
or
Momentarily
Enter a value from 0 (default) to 40 ppt
Save the value
MODE
Momentarily
Saves the new Salinity Compensation
Value
9/11 DirectLine®
DL424/425 Sensor Module User Manual
15
Содержание DirectLine DL424
Страница 15: ...Installation Figure 2 4 Remote Mounting 9 11 DirectLine DL424 425 Sensor Module User Manual 9 ...
Страница 34: ...Spare Parts 28 DirectLine DL424 425 Sensor Module User Manual 9 11 ...
Страница 37: ...Appendix CE Mark Applications 9 11 DirectLine DL424 425 Sensor Module User Manual 31 ...