Problem
Cause
Solution
The sensor filter is blocked.
Replace sensor filter
The detector does not
alarm.
Alarm setpoints set
incorrectly.
Define the alarm setpoint in
Device Configurator.
Alarm setpoints set to zero.
Define the alarm setpoint in
Device Configurator.
Apparatus is in calibration
mode.
Complete the calibration
procedure.
Apparatus is in DC mode.
Stop data communication via
a mobile phone.
Apparatus is in IR
communication.
Stop data communication via
IR Link.
The device alarms without
reason
The sensor is exposed to a
puff of the target gas.
Apparatus is operating
normally. Use caution in
suspected areas. Check the
peak gas exposure reading.
Alarm setpoints are set
incorrectly.
Define the alarm setpoint in
Device Configurator.
Sensors require calibration.
Carry out calibration.
Missing or faulty sensors.
Replace the sensors.
Battery temperature is out
of acceptable range.
Move to lower temperature
ambient to charge the battery.
Battery indicator doesn’t
display when charging.
Battery is depleted.
Charge the battery for 8 hours.
If the battery indicator doesn’t
light after charging, contact
Honeywell
Battery icon flashes in
blue.
Battery is out of required
charging temperature
range.
Move to 0~45°C ambient
temperature.
BW Icon
26
User Manual