3.
In your mobile device, open the Device Configurator app
4.
In your BW Icon:
l
Double press to enter the menu
l
Single press until the BLE icon is displayed
l
Hold press 3 secs to initiate the pairing mode.
5.
In the Device list screen from the Device Configurator app, select the BW Icon Serial
Number to start pairing.
6.
In your Mobile device:
l
Tap on the Menu button
l
Select Bump
l
Follow onscreen instructions.
7. In the mobile device, open the Device Configurator App, and then go to the main menu and
select Bump.
8. Enter the Operator Name, and then Tap Save. The IntelliFlash LED flashes amber to indicate
the calibration process has started.
9. In the Input Gas level screen, check the sensor that you want to calibrate and enter the Span
gas concentration, and then tap OK.
10. Open the cylinder valve by turning the pressure regulator knob clockcounterwise. The Zero
process starts and a message is displayed when succeeded.
11. Follow onscreen instructions to know when to apply gas and when the bump test process is
complete.
12. The process is complete when the results are displayed on your mobile device. You can now
remove the cap by pulling on the tabs.