2. Attach the hose.
3.
In your mobile device, open the Device Configurator app
4.
In your BW Icon:
l
Double press to enter the menu
l
Single press until the BLE icon is displayed
l
Hold press 3 secs to initiate the pairing mode.
5.
In the Device list screen from the Device Configurator app, select the BW Icon Serial
Number to start pairing.
6.
In your Mobile device:
l
Tap on the Menu button
l
Select Remote Calibration
l
Follow onscreen instructions.
7. In the mobile device, open the Device Configurator App, and then go to the main menu and
select Remote Calibration.
8. Enter the Operator Name, and then Tap Save. The IntelliFlash LED flashes amber to indicate
the calibration process has started.
9. In the Input Gas level screen, check the sensor that you want to calibrate and enter the Span
gas concentration, and then tap OK.