26
301W User Manual
511398
Honeywell
5/07
Using the Transmitter
How to Program the Transmitter
The 301W sensors are designed to be maintenance free and do not
require calibration. The following calibration procedures (Functions 7
and 8) are provided as an exceptional troubleshooting measure.
Function 7 - Adjust Sensor Zero
•
Press pushbutton 1
•
Press pushbutton 2, count 8 blinks and release.
•
Begin emitting the calibration gas
•
Press pushbutton 1.
Note: Never calibrate any unit’s Zero with ambiant air. Always use
Nitrogen (N
2
) at the calibration port to calibrate the Zero.
Once this function has been executed, the LEDs will display the result:
Success: Both LEDs blink simultaneously 7 times, for three cycles.
Failure:
LED 1 blinks 7 times, for 3 cycles while LED 2 blinks non-
stop throughout the cycles.
Function 8 - Adjust Span
Allows you to calibrate the sensor’s span. Remember to begin emitting
the calibration gas before starting the calibration.
•
Press pushbutton 1
•
Press pushbutton 2, count 9 blinks and release.
•
Press pushbutton 1.
Once this function has been executed, the LEDs will display the result:
Success: Both LEDs blink simultaneously 8 times, for three cycles.
Failure:
LED 1 blinks 8 times, for 3 cycles while LED 2 blinks non-
stop throughout the cycles.
Содержание 301W
Страница 2: ......
Страница 3: ...Wireless Gas Monitoring Device 301W Installation and User Manual ERP 511398 5 07 ...
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 10: ......
Страница 22: ......
Страница 34: ......
Страница 38: ......
Страница 44: ......
Страница 45: ...Dispositif de détection de gaz sans fils 301W Manuel de l utilisateur ERP 511398 5 07 ...
Страница 46: ......
Страница 48: ......
Страница 50: ......
Страница 52: ......
Страница 64: ......
Страница 72: ......
Страница 86: ......