30
29
TROUBLESHOOTING
If any abnormality arises during use, please check the following points
Note:
If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on the
warranty page.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn on
when the
START/STOP
button
is pushed.
Batteries have run down.
Replace them with two
new AA alkaline batteries.
Battery polarities have
been positioned
incorrectly.
Re-insert the batteries in
the correct positions.
EE measurement error
symbol shown on display
or the blood pressure
value is displayed
excessively low (or high).
Cuff has been placed
incorrectly.
Wrap the cuff properly so
that it is positioned
correctly.
Did you talk or move
during measurement?
Measure again. Keep arm
steady during
measurement.
Shaking of the arm with
the cuff on.
Er error symbol shown on
display - measurement
result exceeds
measurement range.
Did you move during
measurement?
Shaking of the arm with
the cuff on.
Measure again. Do
not move during
measurement.
E1 error symbol shown
on display
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Check cuff connection.
Measure again.
E2 error symbol shown
on display
Inflation pressure
exceeding 300 mmHg.
Turn the unit off, then
measure again.
E3 error symbol shown
on display
Data error.
Remove and reload the
batteries.
CA appears on display
Recalibration alert.
Refer to page 25 for more
information.
Power Source:
Two 1.5V DC (AA) batteries
Measurement
Method:
Oscillometric
Measurement Range:
Pressure: 40~250 millimeters mercury (mmHg)
Pulse: 40~199 beats/minute
Accuracy:
Pressure: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Manual Inflation
Deflation:
Automatic Pressure release valve
Memory Capacity:
60 memories
Auto-shut-off:
1 minute after last button operation
Operation
Environment:
Temperature: 50°F~104°F (10°C ~40°C)
Humidity: 40~85% RH max.
Storage
Environment:
Temperature: 14°F~140°F (-10°C~60°C)
Humidity: 10~90% RH max.
Weight:
.30 lbs/338 grams (without batteries)
Arm Circumference:
Standard size adult arm cuff:
9.5"–14.2" (24 – 36 cm)
Dimensions:
4.50"(L) x 4.17"(W) x 2.25"(H)
112mm(L) x 106mm(W) x 58mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
SPECIFICATIONS
Note:
These specifications are subject to change without notice.