7
Fitting the cuff
1. If the cuff is not assembled at first use, put the end of the cuff which is furthest away from the
tube through the D-ring to form a noose. (Fig.2)
2. Apply the tube to the left side of the device. (Fig.3)
3. Remove tight-fitting clothing from your left upper arm and position the cuff at approximately
2-3 cm above the elbow so that the tube is centered at the middle of the arm. The tube must face
downward towards the hand. (Fig.4)
4. Pull gently at the end of the cuff so that it tightens easily and smoothly around the arm. The cuff
is secured by pulling the velcro fastener towards the cuff.
5. You should be able to fit two fingers between the cuff and your arm. The cuff should not be fitted
any tighter than this. (Fig.5)
6. Relax your body and rest your arm on a table with your palm up so that the cuff is positioned at
the same height as the heart. Make sure that the tube does not bend. (Fig.6)
If it is not possible to fit the cuff on the left arm, it can be placed on the right arm instead. However,
it is important that all measurements are carried out on the same arm. If the measurement is to be
taken on the right arm, the tubel on the cuff should be positioned above the main artery. Press the
arm’s arteries with two fingers at approximately 2 cm above the elbow on the inside of the arm to
find out where the pulse is strongest.
This is the main artery. (Fig.7)
IMPORTANT!
The size of the cuff is instrumental in ensuring accurate measurement results. The cuff has an
appropriate size if the INDEX arrow is placed within the two OK arrows as illustrated below.
If the arrow falls outside this area, another cuff (XL or XS) must be used. (Fig.8)
Important
•
The blood pressure changes with every heartbeat and is consequently changing constantly
24 hours a day.
•
The measurement of blood pressure can be influenced by the user’s position, condition, or other
factors. To obtain the highest degree of accuracy it is recommended that the measurement is
taken 1 hour after workout, bathing, finishing a meal, consuming alcoholic
or caffeine beverages, or after smoking.
•
Before measurement it is recommended that you sit still and relax for 15 minutes. The user
should not be physically tired or exhausted when the measurement is taken.
•
The user must not talk or move arm or hand during measurement.
G
B