| 14
TROUBLESHOOTING
Message
Situation
Solution
Temperature taken in not within Typical human
temperature range.
(34.0°C ~ 43.0°C or 93.2°F ~ 109.4°F).
Make sure the forehead thermometer is for
forehead measurement, not other human
body site.
Measured over the distance: 1-5cm (0.4-2in)
Optimum measurement distance is 1cm
Incorrect test position.
See figure 2 Measuring position and distance.
Subjects forehead hair, Antipyretic stickers, head
with sweat, etc.
Subjects sit quietly 5-10 minutes before
the test.
Some people’s body temperature is lower than the
general population.
The main concern fever temperature
Operating temperature exceeds the range of
specified temperature.
Move to a room within the operating range
wait 30 minutes before taking temperature.
The screen flicker, automatic turn off.
Replace battery or the product has been
damaged, needs repairs.
Battery capacity is too low. Taking
Temperature is not allowed.
Install a new battery
Ambient temperature changes too fast
Wait until the ambient temperature is stable.
1. Power is off.
2. Improper battery installation.
3. The battery is exhausted.
4. Display remains blank.
1. Press ON button again.
2. Check the battery polarity.
3. Replace with a new battery.
4. Contact the retailer or service center.
REPLACING THE BATTERY
1. Open and release battery cover following indicator on the surface of battery cover.
Before changing the battery be sure the thermometer is already turned off.
2. Remove the battery and replace with 2 new type AAA batteries, Make sure to align
properly as indicated inside the battery cover.
3. Slide the battery cover back in until it snaps in place.
Do not dispose of used batteries in household waste. Take them to special local
collection sites.