background image

©

 Home Training Tools Ltd. 2012                                        Page 8 of 8                                         Visit www.homesciencetools.com 

 

 

Poor resolution,  
image not sharp

 

1.

 

The condenser, objective, or eyepiece lenses are dirty. 

Clean the lenses. (See 

“Cleaning,” p. 2.)

 

2.

 

There is too much light. 

Adjust the diaphragm.

  

Spots in field 

1.

 

The condenser, objective, or eyepiece lenses are dirty. 

Clean the lenses. (See 

“Cleaning,” p. 2.)

 

2.

 

The specimen slide is dirty. 

Clean the slide.

  

Uneven 
illumination of 
field 

1.

 

The nosepiece is not indexed properly. 

Move revolving nosepiece until the objective 

lens clicks into position.  

2.

 

The diaphragm is not properly indexed. 

Adjust the diaphragm to the proper level. 

Specifications 

Eyepiece 

Widefield 10x with 18 mm exit pupil and fully coated optics.

Head 

Monocular, 45° inclined head rotates 360°. MI-4100LDH has a dual head: one 30º inclined 
eyepiece, one vertical eyepiece with diopter.

 

MI-4100LBH has a binocular head with dual 

diopters. 

Nosepiece 

3-hole or 4-hole nosepiece is ball-bearing mounted with positive click stops. 

Objectives 

All objectives are DIN achromatic, parfocalled, parcentered, and fully coated. 

 

4x, 0.10 N.A., red ring, 4.5 mm field of view, 40x magnification 

 

10x, 0.25 N.A., yellow ring, 1.8 mm field of view, 100x magnification 

 

40xR, 0.65 N.A., blue ring, 0.45 mm field of view, 400x magnification, retractable 

 

100xR, 1.25 N.A., white ring, 0.18 mm field of view, 1000x magnification, retractable, oil 
immersion (model MI-4100LXL and 4100LBH only) 

Focusing 

Separate low position coarse and fine focusing controls with slip clutch, and tension 
adjustment. 

Focus Rack 

All metal rack-and-pinion focusing with adjustable stage stop.

Stage 

Acid and chemical resistant 110 x 120 mm metal stage with stage clips and predrilled for an 
optional mechanical stage. 

Mechanical 
Stage 

Top-position x-y controls provide precise 60 mm movement on the x-axis and 30 mm 
movement on the y-axis (all models except MI-4100LST). 

Condenser 

Fixed 0.65 N.A. condenser on model MI-4100LST. Adjustable 1.25 N.A. Abbe condenser on 
all other models. 

Diaphragm 

Calibrated 5-hole disc diaphragm on model MI-4100LST. Iris diaphragm with 2 mm to 30 mm 
diameter opening on all other models. 

Illuminator 

Bright LED illuminator with variable intensity control and grounded 110 volt cord.

Warranty 

Home Science Tools warrants this microscope to be free from defects in material and workmanship under normal use and 
service for the life of the instrument. This warranty does not cover light bulbs, batteries, or damage due to misuse, abuse, 
alterations, or accident. Warranty does not cover lenses that have become inoperable due to excessive dirtiness as a 
result of misuse or lack of normal maintenance. Any cameras and software supplied with this microscope are under 
warranty for one year from the date of purchase.  

You will need to return your microscope freight prepaid for warranty service to Home Science Tools, or the repair facility 
we designate. We will repair or replace your microscope at no charge and return it freight prepaid to you. Please call 1-
800-860-6272 to arrange warranty service before returning this instrument.  

Please note that warranties apply only to the original purchaser and are not transferable.

 

Содержание MI-4100LST

Страница 1: ...Home Iris Diaphragm Microscope MI 4100LDH Home Dual head Microscope MI 4100LXL Home LED 1000x Microscope MI 4100LBH Home Binocular Microscope 1 800 860 6272 www homesciencetools com Copyright 2012 by...

Страница 2: ...Adjusting Tension 7 Troubleshooting 7 Specifications 8 Warranty 8 General Microscope Care Unpacking The Home microscopes are shipped in a two part Styrofoam case Keep this case for storage transport...

Страница 3: ...nd down when you turn the focus knobs allowing you to control the distance between the slide and the lens The MI 4100LXL 4100LMS 4100LDH and 4100LBH models have an integrated mechanical stage that all...

Страница 4: ...is for the first time If you do not have a prepared slide place a strand of colored yarn or thread on a blank slide and place a coverslip over it Move the slide until the specimen is under the objecti...

Страница 5: ...m range of motion along the y axis and a 30 mm range of motion along the x axis Filters The MI 4100LXL and 4100LBH models come with a blue filter that can be placed in the filter holder located below...

Страница 6: ...wobble When the head is stable the slide is easier to view 6 To readjust the binocular head loosen the screw turn the head it rotates 360 and then tighten the set screw making sure the head is stable...

Страница 7: ...e to drift down under its own weight and moving the image out of focus The tension adjustment collar is located between the microscope arm and the coarse focus knob on the right side when the stage is...

Страница 8: ...th slip clutch and tension adjustment Focus Rack All metal rack and pinion focusing with adjustable stage stop Stage Acid and chemical resistant 110 x 120 mm metal stage with stage clips and predrille...

Отзывы: