4. If stubborn material has dried on the face or housing of the transducer,
carefully scrub the transducer using water moistened gauze, sponge
or surgeon’s soft bristle brush. Wipe away any material with a water-
moistened soft cloth.
5. Rinse the transducer thoroughly with water. Do not allow water to
access the cable-housing junction.
6. Air dry or dry using a soft dry cloth.
DISINFECTION PROCEDURE
1. Always clean the transducer prior to disinfection. (See steps above.)
2. When using an FDA or CE -cleared (as appropriate) disinfecting
agent, carefully follow the instructions provided by the manufacturer
of the product.
3. For low-level disinfection, the lens and housing of the transducer may
be wiped with any of the
cleaners/low level disinfectants
listed in the associated transducer table.
4. For high-level disinfection, the transducer may be wiped or soaked
using one of the
disinfectants
listed in the associated transducer
table.
5. Air dry or dry using a soft dry cloth.
6. Re-inspect the transducer for any signs of damage such as cracks,
splitting, delamination of the face or cable damage before putting the
transducer back into use.