MyoSure

®

 XL Tissue Removal Device for Fluent

Instructions for Use

Please read all information carefully.

Description

The MyoSure XL Tissue Removal Device for Fluent is a sterile, hand-held tissue removal device used to remove intrauterine tissue. It is connected 

via a flexible drive shaft to the Fluent Fluid Management System. A foot pedal allows the user to control the tissue removal device by turning the 
motor in the Fluent Fluid Management System on and off.
RX ONLY (U.S.) Federal law restricts this device to sale by or on the order of a physician pursuant to 21 CFR 801.109(b)(1).

Indications for Use

The MyoSure XL Tissue Removal Device for Fluent is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove 
tissue such as: Submucous myomas, Endometrial Polyps, and Retained products of conception.

Contraindications

The MyoSure XL Tissue Removal Device for Fluent should not be used with pregnant patients or patients exhibiting pelvic infection, cervical 
malignancies, or previously diagnosed uterine cancer.

Warnings and Precautions

The brief operating instructions in this guide will make the system easier to use. As with any surgical instrument, there are important health and 
safety considerations. These are as follows:

•  Warning: Please consider pre-operative imaging prior to the procedure to assess the patient for evidence of placental invasion of the 

myometrium. In the immediate postpartum phase, removal of retained products of conception (RPOC) in the setting of known or suspected 
placenta acreta, placenta increta or placenta percreta poses a risk of significant and potentially life threatening bleeding.

• Before using the MyoSure XL Tissue Removal Device for Fluent for the first time, please review all available product information.
• Before using the MyoSure XL Tissue Removal Device for Fluent, you should be experienced in hysteroscopic surgery with powered 
   instruments. Healthy uterine tissue can be injured by improper use of the tissue removal device. Use every available means to avoid such    
   injury.
•  The MyoSure XL Tissue Removal Device for Fluent is only compatible with the Fluent Fluid Management System. Use of any other motorized 

power source or fluid management system may fail to operate the device or lead to patient or physician injury.

• If visualization is lost at any point during a procedure, stop cutting immediately.
• Periodic irrigation of the tissue removal device tip is recommended to provide adequate cooling and to prevent accumulation of excised 
   materials in the surgical site.

DANGER: Risk of explosion if used in the presence of flammable anesthetics.
WARNING:  Exercise extreme caution when resecting tissue in patients who have implants that extend into the uterine cavity.

• Do not use the MyoSure XL Tissue Removal Device for Fluent to resect tissue that is adjacent to an implant. When resecting tissue in  patients 
   that have implants, assure that:

• the device’s cutting window is facing away from (i.e., 180° opposite) the implant;
• the visual field is clear; and
• the device’s cutting window is engaged in tissue and is moved away from the implant as tissue resection proceeds.

• In the event an implant becomes entangled with a MyoSure cutter, the following steps are recommended:

• cease cutting immediately;
• kink the device’s outflow tube to prevent a loss of uterine distension;
• disconnect the device’s drive cable from the Fluent Fluid Management System;
• grasp the end of the device drive cable with a hemostat or other clamping device;
• hold the drive cable hub and tissue removal device to prevent twisting;
• open the tissue removal device cutting window by manually twisting the hemostat; and
•  gently pull the device into the hysteroscope to detach the MyoSure XL Tissue Removal Device for Fluent from the implant.

• If this unit is configured as part of a system, the entire system should be tested for compliance with IEC 60601-1-1.

ONLY

Содержание MyoSure XL

Страница 1: ...his document contains proprietary data of Hologic Inc No disclosure reproduction or use of any part thereof may be made except by written permission from Hologic REV RELEASE DATE 15 MAY 2019 Artwork c...

Страница 2: ...t you should be experienced in hysteroscopic surgery with powered instruments Healthy uterine tissue can be injured by improper use of the tissue removal device Use every available means to avoid such...

Страница 3: ...ain distension If this situation occurs tap the foot pedal once or twice to run the tissue removal device the cutting window will then close automatically If it appears that the tissue removal device...

Страница 4: ...ay be affected RF emissions CISPR 11 Class A The FLUENT FLUID MANAGEMENT SYSTEM is suitable for use in all establish ments other than domestic and those directly connected to the public low volt age p...

Страница 5: ...ended separation distance calcu lated from the equation applicable to the frequency of the transmitter Recommended safety distance d 1 2 P for 150 KHz to 80 MHz d 1 2 P for 80 MHz to 800 MHz d 2 3 P f...

Страница 6: ...jects and people Tissue Removal Device for Fluent 50 601XL 50 603XL The MyoSure XL Tissue Removal Device for Fluent is shown in Figure 1 It is a hand held unit which is connected to the Fluent Fluid M...

Страница 7: ...s to the set speed and continues until the foot pedal is released 3 Press the foot pedal and observe the tissue removal device action to verify that the motor runs and that the cutting window is close...

Страница 8: ...pressly stated in the Agreement i Equipment manufactured by Hologic is warranted to the original Customer to perform substantially in accordance with published product specifications for one 1 year st...

Страница 9: ...ox with the returned unit Return used or opened product according to the instructions provided with the Hologic supplied biohazard kit For More Information For technical support or reorder information...

Страница 10: ...xide U S federal law restricts this device to sale by or on the order of a physician Do not resterilize Do not use if package is damaged Hologic MyoSure and associated logos are registered trademarks...

Отзывы: