Cenova Image Analytics Server 3.0 Installation & Service Manual
Chapter 3: Installing the System
Page 34
MAN-05205 Revision 003
3.4.3 Configuring DICOM Destination and Source Devices
The System Settings – DICOM screen lets you configure the server to accept input from
and send results to various DICOM devices, in particular, workstations and PACS
devices. In addition to the initial configuration of DICOM devices, you can also use the
DICOM screen to edit or delete the configurations.
This section provides information on the following:
•
Adding a New DICOM Destination Device
on page 34
•
Configuring DICOM Source Devices
on page 36
•
Understanding the Default Source Device
on page 39
•
on page 40
•
on page 42
Adding a New DICOM Destination Device
Use this procedure to configure the server to send results to a destination device such as
a diagnostic review workstation or PACS.
To add a new DICOM destination device:
1.
Click
DICOM
. The DICOM screen appears.