Affirm Instructions for Use
Chapter 5
—
Troubleshooting
Error Messages
42
Part Number MAN-02866
Table
8:
Affirm
Error
Messages
Error
Message
How
to
Correct
Comm
Error
1.
Select
the
system
status
icon
in
the
taskbar
of
the
Selenia
Dimensions.
2.
Select
the
Clear
All
Faults
options.
This
error
can
occur
when
the
manual
Z
‐
control
is
turned
quickly.
Hardware
Error
1.
Select
the
system
status
icon
in
the
taskbar
of
the
Selenia
Dimensions.
2.
Select
the
Clear
All
Faults
options.
3.
If
the
message
continues
to
display,
contact
Technical
Support.
Selftest
Error
1.
Disconnect
the
Affirm
cable
from
the
Selenia
Dimensions.
2.
Re
‐
connect
the
Affirm
to
the
Selenia
Dimensions.
3.
If
the
message
continues
to
display,
contact
Technical
Support.
Stuck
Switch
Fault
1.
Disconnect
the
Biopsy
Control
Module
from
the
Biopsy
Guidance
Module.
2.
Re
‐
connect
the
Biopsy
Control
Module
to
the
Biopsy
Guidance
Module.
Содержание Affirm MAN-02866
Страница 1: ......
Страница 2: ......
Страница 3: ...Instructions for Use For Software Version 1 6 Part Number MAN 02866 Revision 002 May 2012...
Страница 4: ......
Страница 8: ......
Страница 10: ......
Страница 16: ......
Страница 24: ......
Страница 52: ......
Страница 56: ......
Страница 60: ......
Страница 62: ......