SERVICE
GUIDE
SUPPORT: TOLL-FREE 888-383-DIVE │ FAX 510-729-5115 │ E-MAIL [email protected]
©
2002 DESIGN
(
2012)
DISASSEMBLY
1. Slide the
SILICONE HOSE CLAMP COVER (15, 23)
off of the
HOSE CLAMPS (14, 24)
on the
INHALATION
HOSE (13)
and
EXHALATION HOSE (25)
, where it
mounts to the DSV.
2. Using hose clamp pliers, remove the
HOSE CLAMPS
(14, 24)
from the
INHALATION HOSE (13)
and
EXHA-
LATION HOSE (25) (FIG. 1)
.
3. Gently pull the
INHALATION HOSE (13)
off of the
HOSE INLET HOLDER (17) (Fig. 2)
.
4. Turning counterclockwise, remove both
HOSE NUTS
(16, 22)
from the DSV.
5. Inspect the
HOSE NUT (16, 22)
threads for damage.
Clean and/or replace them as needed.
6. Pull the
HOSE INLET HOLDER (17)
and
HOSE
OUTLET ADAPTER (21)
straight out of the DSV
(Fig.
3)
.
7. With the rounded handle of a screw driver, similar
hand tool, or dowel press the
OUTLET HOLDER (11)
from the DSV
(Fig. 4)
.
8. Inspect the seating surface on the
OUTLET HOLDER
(11)
and
INLET HOLDER (17)
. The surfaces should
be smooth and provide a good sealing surface for the
Mushroom Valves.
9. If the
MUSHROOM VALVES (12, 19)
require replace-
ment, it may be removed by grasping it at the Flange
and pulling it straight out, snipping the Retainer Stem if
necessary. Discard.
10. Remove the
O-RINGS (10, 18, 20)
from the
OUT-
LET HOLDER (11)
,
OUTLET ADAPTER (21)
and
INLET
HOLDER (17)
. Discard.
Fig. 1
Fig. 2
Fig. 3
Fig. 4