Calibration instructions                   

geodyna 6300

Page  7 of 8

Date 2 Oct 2001

Calibration of residual unbalance of main shaft 
and drive pulley using code C84 

Service and test functions become accessible by simultaneously pressing and

holding the C and the balancing mode keys and by rotating the main shaft. 

The following reading appears for one second:

Then the following reading appears:
        
Press the START key to start a measuring run. 

Then the following reading appears:

Calibration is completed and must be stored by entering C90.

 

C84.

C84.      Start

C - -

Содержание geodyna 6300

Страница 1: ...e slider voltage of the diameter potentiometer is not within a range of 15 to 20 Volt turn the potentiometer shaft to bring the voltage to within this range The right hand reading refers to the basic setting of the distance potentiometer If the gauge arm is in left home position and if the slider voltage of the distance potentiometer is not within a range of 15 to 20 Volt turn the potentiometer sh...

Страница 2: ...arm and pull out at least 3 times slowly until it clamps After each clamping hold the gauge arm for at least 1 second in the instantaneous clamping position before repeating this procedure When the gauge arm had been pulled out and clamped 3 times the reading proceeds automatically with step no 7 The following reading appears Pull out the gauge arm at least 7 times with increased constant speed un...

Страница 3: ...ecome accessible by simultaneously pressing and holding the C and the balancing mode keys and by rotating the main shaft The following reading appears Apply the gauge arm on the contact surface of the test rotor and press the C key to store the value The following reading appears Calibration is completed and must be stored by entering C90 C 81 C 1 0 ...

Страница 4: ...e potentiometer shaft to bring the voltage to within this range Press the C key to store the value The following reading appears Apply the gauge head of the width gauge to the outer surface of the adaptor flange and press the C key to store the value The following reading appears Insert the calibration weight supplied with the machine into the adaptor flange so that it protrudes to the right Apply...

Страница 5: ...n function is enabled as soon as the C and balancing mode keys are released Then the following reading appears Press the START key to start a measuring run After the measuring run the following reading appears 2 step no 2 100 suggested calibration weight in grams If necessary press and hold the precision key and rotate the main shaft to set the actual weight Insert a calibration weight of 100 gms ...

Страница 6: ... ambient transducer temperature is read out for 1 second After temperature measurement the following reading appears Remove the 100 gm calibration weight from the test rotor Press the START key in order to proceed with instantaneous electrical compensation of residual adaptor unbalance After the measuring run the following reading appears Insert the calibration weight supplied with the machine at ...

Страница 7: ...unctions become accessible by simultaneously pressing and holding the C and the balancing mode keys and by rotating the main shaft The following reading appears for one second Then the following reading appears Press the START key to start a measuring run Then the following reading appears Calibration is completed and must be stored by entering C90 C84 C84 Start C ...

Страница 8: ...by simultaneously pressing and holding the C and the balancing mode keys and by rotating the main shaft The following reading appears Press the precision key and rotate the main shaft to set 1 The following reading appears Press the C key to complete calibration The following reading appears Press the STOP key to quit the service mode C 90 0 C 90 1 C ...

Отзывы: