- 6 -
3.3 REPROCESSIng PROCEDuRE 3
Procedure 3 can only be applied if the device was used with a bacteria filter, both at the inspiration
port and at the expiration port, and the bacteria filter was changed daily, or according to the manu-
facturer’s instructions. This is the only way to ensure that the device is free of human pathogens.
To reprocess the device, carry out the following steps:
• Change the bacteria filter.
• Disinfect all the parts of the housing and the connections with a suitable agent, e.g., Mikrozid®
Liquid.
• Change the coarse filter and the fine filter and disinfect the surface of the filter cassette. You can
also replace the entire filter cassette with a new one.
4 bloCk Diagrams
4.1 BLOCk DIAgRAM – PnEuMATIC (CARAT I)
Figure 1: Block Diagram – Pneumatic CARAT I
Содержание CARAT I
Страница 1: ...Service Manual Ventilator CARAT I and CARAT II valid for devices of the types 9LV102 and 9LV202 ...
Страница 24: ... 24 10 Exploded Views 10 1 Exploded View C Inner Casing CARAT I ...
Страница 25: ... 25 10 2 Exploded View C Inner Casing CARAT II ...
Страница 26: ... 26 10 3 Exploded View C Inner Casing CARAT II ...
Страница 27: ... 27 10 4 Exploded View A Top Cover CARAT I ...
Страница 28: ... 28 10 5 Exploded View A Top Cover CARAT II ...
Страница 29: ... 29 10 6 Exploded View E Sensor Manifold CARAT I ...
Страница 30: ... 30 10 7 Exploded View E Sensor Manifold CARAT II ...
Страница 31: ... 31 10 8 Exploded View B Lower Cover CARAT rear view ...
Страница 32: ... 32 10 9 Exploded View B Lower Cover CARAT front view ...
Страница 33: ... 33 10 10 Exploded View Assembling CARAT I ...
Страница 34: ... 34 10 11 Exploded View Assembling CARAT I ...
Страница 35: ... 35 10 12 Exploded View Assembling CARAT II ...
Страница 36: ... 36 10 13 Exploded View Assembling CARAT II ...