8.13 CHECk THE ALARMS
You must adhere to the procedure given in chapter “Functional Check” of the User's Manual.
8.14 DATA BACkuP
If it is proposed to change the set parameters during a SRC or maintenance, the contents of the
device memory must be read and saved. If necessary, the data is to be analyzed. no maintenance
is necessary to the PC software.
8.15 STAnDARD VALuES THE FACTORy SETTIng
Parameter values in accordance with the factory setting (through a PC or direct on the device)
general Parameters
Mode
PCV
Display I:E
I:E
Display Pressure Units
hPa
Display Screen
STD
Safety Lock
Manual
Safety Lock (Machine)
OFF
Volume
50 %
Pressure Deviation
3.0 hPa
Max. Inspiration time
4 s
Leakage
OFF
Max. Oxygen
OFF
Min. Oxygen
OFF
PIn code
1302
Device Language
Eng
Parameter PCv
IPAP
20.0 hPa
PEEP
5.0 hPa
Frequency
12 bpm
I:E
1 : 1.5
Chute
3
Inspiration trigger
OFF
Min. Volume
OFF
Additional pressure
5.0 hPa (not active, Min. volume is set to OFF)
Max. Volume
1.00 l
Min. Volume
0.20 l
Max. Frequency
30 bpm
Содержание CARAT I
Страница 1: ...Service Manual Ventilator CARAT I and CARAT II valid for devices of the types 9LV102 and 9LV202 ...
Страница 24: ... 24 10 Exploded Views 10 1 Exploded View C Inner Casing CARAT I ...
Страница 25: ... 25 10 2 Exploded View C Inner Casing CARAT II ...
Страница 26: ... 26 10 3 Exploded View C Inner Casing CARAT II ...
Страница 27: ... 27 10 4 Exploded View A Top Cover CARAT I ...
Страница 28: ... 28 10 5 Exploded View A Top Cover CARAT II ...
Страница 29: ... 29 10 6 Exploded View E Sensor Manifold CARAT I ...
Страница 30: ... 30 10 7 Exploded View E Sensor Manifold CARAT II ...
Страница 31: ... 31 10 8 Exploded View B Lower Cover CARAT rear view ...
Страница 32: ... 32 10 9 Exploded View B Lower Cover CARAT front view ...
Страница 33: ... 33 10 10 Exploded View Assembling CARAT I ...
Страница 34: ... 34 10 11 Exploded View Assembling CARAT I ...
Страница 35: ... 35 10 12 Exploded View Assembling CARAT II ...
Страница 36: ... 36 10 13 Exploded View Assembling CARAT II ...