Hoefer PS200HC Скачать руководство пользователя страница 18

Connections

1

Connect the mains power cord. Select the appropriate 
mains power cord for your region, and connect it to 
the power cord receptacle on the rear panel and to a 
suitable grounded AC power outlet. Place the power 
supply so that the fan on the back panel and the 
vents on the bottom are not obstructed.

2

Connect the HV power leads. Connect the apparatus 
to the power supply by plugging the lead connectors 
into one pair of the recessed output jacks (Fig 4).

3

Turn on the main power switch on the rear panel. The 
front panel should be illuminated.

 p8

Fig 4. The output jacks are 
recessed and color coded.

black jacks

red jacks

Содержание PS200HC

Страница 1: ...user manual um PS200HC IM Rev C0 07 12 PS200HC Power Supply...

Страница 2: ...formation viii Power supply function and description 1 Unpacking 3 Specifications 4 Parameter overview 5 Power supply operation 7 Connections 8 Operation 9 Power Failure Detection PFD 12 Troubleshooti...

Страница 3: ...it n pou it v laborato i pouze Pouze p slu enstv a sti schv len nebo poskyt nut ch Hoefer Inc mohou b t pou ity pro provoz dr bu a dr b tohoto v robku Pozor Proto e tento n stroj m e vyvinout dostate...

Страница 4: ...spanning zou in overeenstemming met EN61010 1 2001 moeten zijn Houd het instrument zo droge en schone zoals mogelijk Bij Wis regelmatig met een zacht temper doek Verhuur het instrument droogt volledig...

Страница 5: ...sr stung zur Verf gung gestellte Schutz verschlechtert werden Dieses Instrument wird f r den Innenlaborgebrauch nur daf r entworfen Nur Zus tze und Teile genehmigten oder lieferten durch Hoefer Inc ka...

Страница 6: ...ig sjokk m bli vd bekymring i dets drift Dette instrumentet er utformet i samsvar med EN61010 1 2001 elektrisk sikkerhetsstandard Likevel burde bli brukt det bare av skikkelig utdannede oper at rer Le...

Страница 7: ...o h mido macio Deixe o instrumento secar completamente antes de uso N o opera o instrumento em humidade extrema acima de 80 Evite condensa o deixando o equili brate de unidade a temperatura ambiental...

Страница 8: ...as bara av riktigt utbildade operat rer L s denna hela handbok f re anv nda instrumentet och anv nd bara enligt under visningarna Instrumentet m ste alltid anv ndas med jorden blyet av kraften repet r...

Страница 9: ...e Kontakt mit einem autorisierten Beauftragten des Herstellers auf um Informationen hinsichtlich der Entsorgung Ihres Ger tes zu erhalten Questo simbolo indica che i rifiuti derivanti da apparecchiatu...

Страница 10: ...CE certification This product meets the requirements of applicable CE directives A copy of the corresponding Declaration of Conformity is available on request The CE symbol and corresponding declarati...

Страница 11: ...current status Param eters are set by pressing UP DOWN arrow buttons in combination with a SELECT button Four pairs of output jacks accept 4 mm plugs which operate in parallel The output jacks are col...

Страница 12: ...p2 Fig 1 Front view of PS200HC Power Supply 3 digit display UP DOWN arrows SELECT display highlighted parameter CONSTANT V or A toggle START STOP polarity Output jacks polarity...

Страница 13: ...er Inc sales office Inspect all components for damage that may have occurred while the unit was in transit If any part appears damaged contact the carrier immediately Be sure to keep all packing mater...

Страница 14: ...lel Output connector 4 mm plugs Timer 1 999 min 1 min increment Environmental operating Indoor use 4 40 C conditions Relative humidity 0 80 noncondensing Altitude 2000 m Pollution degree 2 Installatio...

Страница 15: ...d or power requirement does not exceed the power supply rating Crossover occurs when the second parameter limit is reached as system resistance changes Fig 3 At this point the second parameter becomes...

Страница 16: ...ower When this occurs both the constant V and the constant A lights will be lit at the same time p6 current Voltage limit Crossover from constant A to constant V Current limit resistance increasing ov...

Страница 17: ...ng or discon necting the voltage leads 2 Keep fingers away from metal components when connecting or disconnecting the voltage leads p7 WARNING Use only one hand when making or breaking a connection to...

Страница 18: ...e power supply so that the fan on the back panel and the vents on the bottom are not obstructed 2 Connect the HV power leads Connect the apparatus to the power supply by plugging the lead connectors i...

Страница 19: ...s to set it to the desired value The power supply can then operate in Crossover mode as explained on page 5 4 Set the timer optional Press the SELECT button to move the green indicator light to under...

Страница 20: ...ng the CONSTANT button will reset the power supply Example Set up conditions for running a standard transfer tank e g TE62 at a constant current of 1 0 A 1 Press the CONSTANT button to position the in...

Страница 21: ...his value down to 30 V 4 Press the SELECT button to position the indicator light under the Clock symbol If desired use the UP button to set the timer to a desired duration If the run is continuous set...

Страница 22: ...The time shown at the end of the run is the time the gel ran since power was restored not the total time of the run The default settings for Power Failure Detection are Run Time Setting PFD Code Time...

Страница 23: ...Check buffer volumes Check all voltage leads Check the fit of power supply adapters if used The output current may have fallen below the 01 A limit of the power supply Press any button to clear the er...

Страница 24: ...el Use a screwdriver to pry out the fuse compartment There are two fuses in the compartment Check and replace any blown fuses with a fuse of the same size and ratings FUSES 2 250 V 5 0 A 5 20 mm fast...

Страница 25: ...l Hoefer Inc for advice before using Technical service and repair Hoefer Inc offers complete technical support for all of our products If you have any ques tions about how to use this product or would...

Страница 26: ...lliston MA 01746 Toll Free 1 800 227 4750 Phone 1 508 893 8999 Fax 1 508 893 0176 E mail support hoeferinc com Web www hoeferinc com Hoefer is a registered trademark of Hoefer Inc 2012 Hoefer Inc All...

Отзывы: