RXW TEROS-12 Soil Moisture EC Temp Sensor (RXW-T12-xxx) Manual 

1-800-LOGGERS 

www.onsetcomp.com 

shows all symbols illuminated on the LCD screen followed by 

definitions of each symbol in the table. 

 

LCD Symbol  Description 

 

The battery indicator shows the approximate battery 

charge remaining. 

 

This is a signal strength indicator. The more bars, the 

stronger the signal between motes. If there is no x 

icon next to the signal strength indicator, then the 

mote is part of a HOBOnet Wireless Sensor Network. 

 

An empty signal strength icon plus the x icon 

indicates that the mote is not currently part of a 

network. See 

Adding the Mote to the HOBOnet 

Wireless Sensor Network 

for details on how to add a 

mote to the network. 

 

When the mote is in the process of joining a 

network, the signal strength icon will blink and then 

the bars in the icon will cycle from left to right. The x 

icon will blink during the last step in the network 

registration process (see 

Adding the Mote to the 

HOBOnet Wireless Sensor 

Network for details). 

 

This indicates a problem with the sensor itself (the 

mote is operational). Check the sensor and make any 

adjustments to it as needed. Contact Onset 

Technical Support if the problem persists. 

Sensor Components and Operation 

The sensor measures soil moisture, temperature, and EC of soil 

using stainless steel needles. As shown in the following 

diagram, the sensor measures soil moisture between Needle 1 

and Needle 2 and measures EC between Needle 2 and Needle 

3. Temperature is measured with an embedded thermistor. 

 

A ferrite core positioned on the sensor cable 7.6 cm (3 

inches) away from the sensor head is used to isolate the 

sensor from any interferences in the system. This 

mitigates any potential noise from the system on the 

measured sensor data. 

Important:

 Do not attach anything to the section of the 

cable between the sensor head and the ferrite core as 

this may influence the measurements. 

The VWC measurement sensitivity is contained within a 1,010 

mL volume roughly depicted in the following VFW volume of 

influence figure. 

 

 

 

Adding the Mote to the HOBOnet Wireless 

Sensor Network 

The mote must join a HOBOnet Wireless Sensor Network 

before it can begin measuring soil moisture and transmitting 

data. This requires accessing the station and the mote at the 

same time so it is recommended that you complete these steps 

before deploying the mote. 

Important: 

If you are setting up a new station, follow the 

instructions in the station quick start before setting up this 

mote (go to www.onsetcomp.com/support/manuals/24380-

man-rx2105-rx2106-qsg for RX2105 and RX2106 stations or go 

to www.onsetcomp.com/support/manuals/18254-MAN-QSG-

RX3000 for RX3000 stations).

 

To add a mote to the network: 
1.

 

If the LCD is blank on the station, press any button to wake 

it up. 

2.

 

Press the Select button once (which shows the number of 

smart sensors installed) and then press it again to switch to 

the module with the manager (module 2 on RX2105 or 

RX2106 stations). 

 

 

 

 

 

 

 

Press this button to view the module 

Needle 3 

Needle 2 

Needle 1 

Ferrite core 

Thermistor in Needle 2 

EC 

VWC 

 

WARNING:

 Risk of 

personal injury is 

present. Sensor needle 

points are extremely 

sharp and will puncture 

skin. Handle with care. 

9.3 cm 

7.5 cm 

Содержание RXW TEROS-12

Страница 1: ...m Wireless Mote Operating Temperature Range Sensor 40 to 60 C 40 to 140 F Mote 25 to 60 C 13 to 140 F with rechargeable batteries 40 to 70 C 40 to 158 F with lithium batteries Radio Power 12 6 mW 11...

Страница 2: ...he mote batteries see Deploying and Mounting Sensor Cable This is the cable that connects the mote to the sensor Eyelet Use this eyelet to attach a 3 16 inch padlock to the mote for security Latch Use...

Страница 3: ...an embedded thermistor A ferrite core positioned on the sensor cable 7 6 cm 3 inches away from the sensor head is used to isolate the sensor from any interferences in the system This mitigates any pot...

Страница 4: ...HOBOlink and report the mote as missing Once the mote is back online any logged data will be uploaded the next time the station connects to HOBOlink See the HOBOlink Help for details on how to change...

Страница 5: ...are contaminated with oils from contact with skin or another source follow the instructions in the Maintenance section for cleaning the needles before installing the sensor Avoid any metal located bet...

Страница 6: ...load the file to your computer 5 In HOBOlink go to Devices then RX Devices and click your station name 6 On the station page click Overview and scroll down to Device Information 7 Click the Wireless t...

Страница 7: ...ll the new ones observing polarity 3 Push in the side tab of the solar panel cable connector and pull the connector out of the cable port 4 Place the connector in the slot on the inside of the mote do...

Страница 8: ...uding interference that may cause undesired operation FCC Caution Any changes or modifications not expressly approved by the party responsible for compliance could void the user s authority to operate...