11
BA 24 (800012900), remove the socket screw in the block with the enclosed
screwdriver. Screw the enclosed threaded pin with isolation knob into the
anti-condensation plate. Place the anti-condensation plate on the block and
screw the threaded pin in deeper so that it takes hold of the bolt of the sys-
tem and with that tightly attaches block and anti-condensation to the system.
Troubleshooting
The display remains blank
Please check if the mains switch on the back side is switched on. If it is
switched on but voltage is present at the socket, check the micro-fuse and
replace if required (IEC 127-2/III, 250 V, 2 A time-lag). This fuse - and a spare
fuse - are located in the IEC-bushing (in which the connection cable is
plugged). The fuse box can be pulled out with a screwdriver.
There are excessive temperature fluctuations
Check the seat of the changeable block by pulling it up. If it is too loose or if
the block exhibits roughness, e.g. contamination, the heat / coolness is not
transferred correctly.
Information on the capacitive touch display
The touch display reacts on finger pressure, even when wearing thin latex
gloves, but not on stylos. Due to its glass surface, the front panel is insensitive
to dirt, chemicals and mechanical damage. Please avoid scratches in the coat-
ing since they can cause malfunctions. Please take note of the following clean-
ing instructions.