Operation / Maintenance
9.1
Troubleshooting
If the device behaves unusual or erratic, please switch off the supply voltage and check the wiring of
the load and that of the control and monitoring signals. Check the connected load and the signal
source. Check the interlock circuit.
During operation various parameters are monitored and analyzed. In case of a fault the condition is
shown via LEDs on the front panel of the amplifier. Press the
SDN RST
push button to reset the error
condition.
Do not attempt to locate any faults within the device. This can be dangerous to life due to the high
voltage used in the device. In such a case, please return the device to the manufacturer after
consultation.
Symptom Possible
causes
Device will not turn on;
POWER
switch will not light up
-
No supply voltage present
-
Fuses F1, F2 in the IEC inlet defective
HV OFF
palm button will not light up
-
HV OFF
palm button is in
off
position (pushed in)
-
Interlock circuit is open
Output signal is distorted
-
Load capacitance too large for the desired slew rate
(see section 3.8 for load conditions)
Amplifier switches off, the
OVERLOAD
indicator lights up
-
Dynamic load current exceeded (see section 3.8 for load
conditions)
-
Static load current exceeded (see section 3.8 for load
conditions)
Amplifier switches off, the
OVERTEMP
indicator lights up
-
Poor ventilation
-
Total load too large (see section 3.8 for load conditions)