-35-
MN1-1162 Rev. 21
High-level disinfection
General name
Trade name
Manufacturer
Hydrogen peroxide
PERASAFE
TM
*
Practical liquid 1.62W/V%
ANTEC INTERNATIONAL
Peracetic acid
Acecide
®
*
Solution 6%
Saraya Co., Ltd.
Glutaraldehyde
WAVICIDE
®
-01 *
Solution 2.65%
Medical Chemical Corporation
Glutaraldehyde
STERANIOS *
Solution 2.0%
Laboratoires ANIOS
Glutaral
Cidex plus
®
Solution 3.4%
ADVANCED STERILIZATION PRODUCTS
®
A Johnson & Johnson company
Division of Ethicon, Inc.
Note: * indicates that the marked disinfectant is not applicable in Canada.
Warning
After disinfection, thoroughly rinse the probe with deionized water.
Residual disinfectant can cause an adverse reaction to the operator or patient.