-18-
MN1-5575 Rev. 11
3-1-4. Probe insertion check
With the protect tube locked to the trocar outer sheath, insert the probe into the protect tube and make
sure the probe can be smoothly inserted/removed.
3-1-5. Verification of operation
Connect to the ultrasound diagnostic instrument by following the instructions in section 4-2, “Connecting
to the ultrasound diagnostic instrument" and check that the selected probe match the linear display and
the displayed frequency and check the image for errors.
Remarks
For details on the displayed screens, see the documentation supplied with the ultrasound diagnostic
instrument.
If the probe is operated in still air, brightness on the top of the image may be non uniform, but this does
not affect the performance of the probe.
Caution
Do not use the probe if the selected probe and image do not match the frequency.
An incorrect acoustic output can result in burns or other injuries to the patient. Contact one of our
offices and/or distributor's offices listed on the back cover.
3-1-6. Verification of cleaning, disinfection and sterilization
Verify that cleaning, disinfection and sterilization are conducted according to the intended use.