5 - 45
Code Error
Message
Countermeasure
1690 The instruction manual does not
exist.
Install the instruction manual.
This message appears if the TM3000 manual (in
PDF format) is not installed on the PC,
preventing the display of the manual.
Install the TM3000 manual from the supplied
Manual CD.
1691 The instruction manual cannot
be displayed.
Confirm if the PDF file browser
has been installed.
This message appears if the PDF file read
software (Adobe Acrobat Reader) not installed
on the PC, preventing the display of the manual.
Install the PDF file read software (Adobe Acrobat
Reader) from the supplied Manual CD.
1706 Observation screen size is
changed.
Do you save this setting?
This confirmation message appears when the
[Close] button is clicked without clicking the
[SET] button after changing the observation
screen size.
Clicking the [OK] button changes the size and
closes the [Setup] screen. Clicking the [Cancel]
button closes the [Setup] screen without
changing the size.
1707 The input value of the
observation screen size is
unacceptable.
This size returns to the previous
registered value.
This message appears when the observation
screen size that was entered is out of range.
Clicking the [OK] button resets the observation
screen size to the currently stored settings.
1720 Re-starting up the program is
required for changing the
language.
Do you exit the application?
This message appears when the modification of
display language is selected.
Clicking the [OK] button applies the change in
display language and starts the termination
processing. Clicking the [Cancel] button cancels
the modification of display language.
1721 Re-starting up the program is
required for changing the option
setting.
Do you exit the application?
This message appears when option settings are
modified.
Clicking the [OK] button applied the change in
option settings and starts the termination
processing. Clicking the [Cancel] cancels the
modification of option settings.
1735 It is time to maintain the Turbo
Molecular Pump.
It is recommended to replace it
with a new one.
This message appears when the turbo molecular
pump run time has reached a specified limit (720
hours).
Click the [OK] button to dismiss the message.