3-2. Start up check of the puncture adapter
3-2-1. Visual check
Check the punctute adapter for any of the abnormal conditions below.
•
Abnormalities seen in visual such as deformation, cracks, abnormal gaps, damage, foreign matter adhering,
severe discoloration.
3-2-2. Mechanical inspection of the puncture adapter
Check that the puncture adapter mechanism while attached to the probe.
• The screws (incl. Fixing knob, Needle groove setting knob and Lever locking knob) have no abnormalities
such as looseness, backlash and immobility.
• When the release lever is pressed with the lever locking knob loosen, the puncture needle can be removed
from the puncture adapter.
• The release lever does not move when the lever locking knob is tightened securely.
The puncture adapter is firmly attached to the probe.
• A puncturing needle that passes through the needle groove moves smoothly in the puncturing direction.
Remark:See section 4-2 “Attaching of the puncture adapter”
3-2-3. Verification of operation
When puncturing under the ultrasonic guide, for safety reason, it is also recommended that you have a full
understanding of ultrasound diagnostic characteristics and conduct practice beforehand using a tub or similar
object.