9 Maintenance
9.3
Maintenance of the spark stand
9.3.1
Disassembling and cleaning the spark stand
When the spark stand must be cleaned, an according message is displayed on the
screen.
Information
We recommend the immediate cleaning of the spark stand when this message is
displayed.
You could skip the message clicking "Remind later", however, that is not recommended.
The cleaning interval can be set in the "Options" window (see software manual).
Proceed as follows:
1.
Switch the instrument off (chapter 9.1, page 39).
Opening the spark
stand
2.
Loosen the 4 holding clamps (1, fig. 9-1) on the spark stand plate (2).
3.
Lift the spark stand plate vertically to the top.
Fig. 9-1
Opening the spark stand plate
Cleaning the spark
stand
4.
Brush the condensate off the surface below the spark stand plate and around the
electrode using the brush from the accessory case.
There is a crush hazard when closing the holding clamps on the spark stand
plate!
•
Close the holding clamps from above.
5.
Replace the spark stand plate on the spark stand and close the holding clamps.
1
2
1
Содержание OE700 Series
Страница 1: ...Operating manual 10022080 Ver 1 0GB 2019 08 14 OE700 series ...
Страница 4: ......
Страница 8: ...Table of figures VI Hitachi High Tech Analytical Science ...
Страница 9: ......
Страница 10: ...Table of figures VIII Hitachi High Tech Analytical Science ...
Страница 14: ......
Страница 20: ......
Страница 26: ......
Страница 36: ......
Страница 46: ......
Страница 48: ......
Страница 62: ......
Страница 66: ......
Страница 68: ......
Страница 70: ......
Страница 72: ......
Страница 74: ......
Страница 76: ......