7 Quick start
►
34 Hitachi High-Tech Analytical Science
7.2.3
"Program" recalibration
Start "Program" recalibration
Select "Recalibration" in analysis
window
Select "Recalibrate Program"
option
Spark desired sample
3 to 4 times minimum
(avoid sparking the same burn
spot)
Continue with "Analysing"
Confirm result
Should one sparking
deviate excessively:
remove by double-
clicking the column
and clicking
"Delete".
OK
Check results