6 Start-up
6
Start-up
The initial start-up of the instrument is carried out by customer service employees of
Hitachi High-Tech Analytical Science or an authorised representative and includes a
complete functional test of the instrument as well as operator training.
If you put the instrument back into operation after a shutdown, please proceed as
follows:
Information
The following description of putting into service assumes that you are familiar with
operating the software and the instrument.
Prerequisite
•
The installation of the instrument has been carried out (chapter 5.4, page 19).
•
Optical transparency has been established (chapter 5.5, page 20).
•
The supplied recalibration samples are available and have been prepared well (i.e.
freshly ground, fat-free sample surface).
Recalibration
1.
Switch on the computer and wait until WINDOWS™ has been completely started
up.
2.
Open the SpArcfire spectrometer software double-clicking the SpArcfire
icon
. The start screen is displayed.
3.
Switch on the spectrometer at the on/off switch (1, fig. 6-1).
Fig. 6-1
On/off switch
4.
Select "Analysis" in the start screen.
5.
Select a matrix and analysis program and confirm your selection with "OK". The
analysis screen is displayed.
6.
Click "Recalibration" to recalibrate the instrument.
7.
Select the "Recalibrate Matrix" option. The name of the first recalibration sample
(e.g. RE12) is displayed in the centre of the screen.
1
Содержание OE700 Series
Страница 1: ...Operating manual 10022080 Ver 1 0GB 2019 08 14 OE700 series ...
Страница 4: ......
Страница 8: ...Table of figures VI Hitachi High Tech Analytical Science ...
Страница 9: ......
Страница 10: ...Table of figures VIII Hitachi High Tech Analytical Science ...
Страница 14: ......
Страница 20: ......
Страница 26: ......
Страница 36: ......
Страница 46: ......
Страница 48: ......
Страница 62: ......
Страница 66: ......
Страница 68: ......
Страница 70: ......
Страница 72: ......
Страница 74: ......
Страница 76: ......