WARNING
5.7
Sterilization
The
probe
and
accessory
can
be
sterilized
using
either
ethylen
oxide
gas
(EtO)
sterilization
or
plasma
sterilization
(see
table
in
the
next
page).
Follow
the
manufacturer
ʹ
s
instructions
of
the
sterilizer
regarding
usage,
temperature
and
sterilization
‐
time.
The
sterilization
method
and
operating
conditions
are
as
follows
Sterilization
Method
Condition
Plasma
Sterilization:
STERRAD
®
50,
100S
or
200
(*)
Short
Cycle
Plasma
Sterilization:
Sterrad
®
NX
or
100NX
(*)
Standard
cycle
ETO
Sterilization
Gas
Type:
10%
EO/
90%
HCFC
Temperature:
50
‐
55
°
C
Exposure
Time:
More
than
120
minutes
Pressurization:
162
‐
200kPa
Depressurization:
13
‐
8kPa
Relative
humidity:
40
‐
90%
Aeration
is
minimum
12
hours
*
STERRAD
®
systems
are
manufactured
by
ʺ
Johnson
&
Johnson
ʺ
Before
performing
sterilization,
check
that
the
operation
data
of
sterilizer
are
in
conjunction
with
min.
and
max.
data
applicable
for
the
probe
and
the
Attachment.
Do
not
sterilize
the
probe
and
the
Attachment
by
Steam
Autoclaving.
If
you
autoclave
them,
they
suffer
serious
damage
and
will
not
be
functional.
Sterilization