Peak range
N•m
0.015 - 1.000
0.15 - 10.00
N•cm
0.15 - 10.00
1.5 - 100.0
15 - 1000
N•mm
1.5 - 100.0
Ibf•in
0.015 - 0.90
0.15 - 9.00
1.5 - 90.0
(kgf•cm)
(0.15 - 10.00)
(1.5 - 100.0)
Accuracy*
within ±0.5% (at full scale)
Power
fi
ve 1.2 V NiMH cells 1,000 mA
Recharge time
up to 6 hours
Running time on full charge
8 hours
(continuous operation)
Battery life
About 300 chargings
Dimensions
See
fi
gure
Weight
1.8
kg
Battery charger
Runs on 100-240 V AC power (50 or 60 Hz)
Output 7.3 - 8 V DC (0.5 A)
Model HP-1
HP-10
HP-100
■
Speci
fi
cations
The display says "
LOBAT
"
There is no display
The displayed value does
not return to zero
The displayed values do
not stop changing
Unable to recharge the
battery
The "
LOBAT
" display
reappears after battery
recharging is complete
T h e d i s p l a y s h o w s a
numerical value when it
shouldn't
Problem
Causes
Resolving the problem
Recharge the battery as explained on page 12. If this
does not resolve the problem, contact HIOS or your
HIOS dealer.
Turn the power switch off, conduct a full battery re-
charge (6 hours, no more), then turn the power switch
on and see if there is a display. If there is not, contact
HIOS or your HIOS dealer.
Conduct zero adjustment.
Set to "
PEAK
."
Check that the plug is connected properly.
Connect to the battery charger socket.
If the above does not remedy the problem, contact
HIOS or your HIOS dealer.
Send to HIOS for repair.
Recharge the battery for no more than 6 hours.
Press
【
RESET
】
switch to clear the display.
Insuf
fi
cient battery charge
This may be due to long-term use of the
meter or natural battery discharge over
time.
Unable to conduct zero adjustment.
The
【
MODE
】
switch is set to "
TRACK
"
The battery charger plug may not be en-
gaged all the way into its socket.
The connection may be to the wrong terminal.
The battery may have reached the end of
its operational life.
The battery charge may be insuf
fi
cient.
This may be due to noise occurring from
the power source, etc.
■
Troubleshooting
Before you conclude the torque meter is broken, go through the following troubleshooting guide. If, by fol-
lowing the instructions, you are unable to resolve the dif
fi
culty, please contact HIOS Corporation or your
HIOS dealer.
• Never exceed the maximum torque tolerance level for the instrument.
• Because battery life depends on operating conditions, this cannot be guaranteed.
- 14 -