①
②
③
④
⑤
■
Fidaptor parts
HIOS 4 mm
TF5G-0010-4
HP-100
HIOS 5 mm
TF5G-0010
Hex
shank
TF6XG-0010
(
fi
ts 5 mm and 6.35 mm sockets)
HP-10
HIOS 4 mm
TF4G-0010
HP-100
TF5-0020
HP-10
CL65-0250
HP-100
TF5G-0030
HP-10
TF4-0030
Black spring (0.5 - 3N•m)
TF5-0051
Yellow spring (0.15 - 0.6N•m)
CL65-0040
HP-100
TF5G-0050
HP-10
TF4G-0050
①
Threaded shaft
Name of part
Type
Part number
②
Bearing set
③
Bearing housing
④
Spring
⑤
Plate nut
Other parts (available for separate purchase)
Item
Driver
Shaft
Measuring
Range
Part No.
Note
Fidaptor
Hex. shank
6.35 mm
subtense
3
−
9N•m
TF6U-Z
*1
For HIOS torque drivers with lever-
start.
Fidaptor
Hex. shank
6.35 mm
subtense
3
−
9N•m
TF6UP-Z
*2
For HIOS CL-9000 or high torque
derives with both lever-start and
push-to-start.
Standard accessory
set for HP-100
TF5G-SET See the Manual p.4 of Standard
accessories (without battery charger).
The safety cover provides protection against the possibility of shaft breakage due to
wearing or metal fatigue, but for an added measure of safety it is recommended that
the user follow the above guide to parts replacement.
Guide to when it's time to replace Fidaptor parts
① , ②
:
Threaded shaft and bearing set: After 2,500 strokes
①
:
Threaded shaft only: When it becomes bent or its threads are worn
④
:
Spring: Replace together with threaded shaft after 2,500 strokes
⑤
:
Plate nut
: After 5,000 strokes
Note
6.
See that each measuring cycle lasts 5 seconds
or more
to minimize wearing of parts.
7. Never attach the Fidaptor unless you intend to
use it immediately. When it is not in use, always
remove it from the torque tool and remove the
tension from the spring.
8. Always be sure that the Fidaptor is properly at-
tached to the torsion tool.
9. Do not use the Fidaptor if it has been bent or in
some way modi
fi
ed.
* Use the grease supplied with the torque meter to
apply to parts as necessary. This grease can be
purchased separately (part number TF-G3).
Maintenance and inspection of the Fidaptor
1. Apply the grease periodically to parts
①
,
②
and
⑤
in the
fi
gure below.
2. Always inspect the parts of the Fidaptor before
use.
(1) Is there grease on parts
①
,
②
and
⑤
?
(2) Have the threads of the threaded shaft be-
come worn?
(3) Is there dirt or foreign material on the
threaded shaft?
3. The Fidaptor must eventually be replaced.
It
should be periodically inspected and replaced
when necessary.