11
Range
9660, 9695-03
9661
500.00A
-
±2.3% rdg. ±0.31% f.s.
100.00A
±2.3% rdg. ±0.32% f.s.
±2.3% rdg. ±0.35% f.s.
50.000A
±2.3% rdg. ±0.34% f.s.
±2.3% rdg. ±0.4% f.s.
10.000A
±2.3% rdg. ±0.5% f.s.
±2.3% rdg. ±0.8% f.s.
5.0000A
±2.3% rdg. ±0.7% f.s.
±2.3% rdg. ±1.3% f.s.
Range
9669
1.0000kA
±3% rdg. ±0.31% f.s.
200.00A
±3% rdg. ±0.35% f.s.
100.00A
±3% rdg. ±0.4% f.s.
Range
9694
9695-02
50.000A
-
±2.3% rdg. ±0.32% f.s.
10.000A
-
±2.3% rdg. ±0.4% f.s.
5.0000A
±2.3% rdg. ±0.32% f.s.
±2.3% rdg. ±0.5% f.s.
1.0000A
±2.3% rdg. ±0.4% f.s.
±2.3% rdg. ±1.3% f.s.
500.00mA
±2.3% rdg. ±0.5% f.s.
±2.3% rdg. ±2.3% f.s.
Range
CT9667-01, -02, -03 5.000kA
range
CT9667-01, -02, -03 500A
range
5.0000kA
±4% rdg. ±0.6% f.s.
-
1.0000kA
±4% rdg. ±1.8% f.s.
-
500.00A
±4% rdg. ±3.3% f.s.
±4% rdg. ±0.6% f.s.
100.00A
-
±4% rdg. ±1.8% f.s.
50.000A
-
±4% rdg. ±3.3% f.s.
CLAMP ON SENSOR 9660 / 9695-03 / 9661
*2
Voltage Connection
Current
9661only
5.0000A
10.000A
50.000A
100.00A
500.00A
400.0V
1P2W
2.0000kW 4.0000kW 20.000kW 40.000kW 200.00kW
1P3W
1P3W1U
3P3W2M
3P3W3M
4.0000kW 8.0000kW 40.000kW 80.000kW 400.00kW
3P4W
6.0000kW 12.000kW 60.000kW 120.00kW 600.00kW
CLAMP ON SENSOR
9669
Voltage Connection
Current
100.00A
200.00A
1.0000kA
400.0V
1P2W
40.000kW
80.000kW
400.00kW
1P3W
1P3W1U
3P3W2M
3P3W3M
80.000kW
160.00kW
800.00kW
3P4W
120.00kW
240.00kW
1.2000MW
CLAMP ON SENSOR 9694 / 9695-02
*1
Voltage Connection
Current
500.00mA 1.0000A
5.0000A
10.000A
50.000A
400.0V
1P2W
200.00W
400.00W 2.0000kW 4.0000kW 20.000kW
1P3W
1P3W1U
3P3W2M
3P3W3M
400.00W
800.00W 4.0000kW 8.0000kW 40.000kW
3P4W
600.00W 1.2000kW 6.0000kW 12.000kW 60.000kW
*1
For the 9694 sensor, the range of guaranteed accuracy is from 500 mA to 5 A,
and for the 9695-02, from 500 mA to 50 A.
*2
For the 9660 and 9695-03 sensors, the range of guaranteed accuracy is from 5 A to 100 A
and for the 9661, from 5 A to 500 A.
Leak current: CLAMP ON LEAK SENSOR 9657-10, 9675
Range 50.000mA / 100.00mA / 500.00mA / 1.0000A / 5.0000A
Apparent power
±1 dgt. for the calculation obtained from each measurement value
Reactive power
Fundamental waveform calculations
±2.0% rdg. ±3.0% f.s. + clamp-on sensor accuracy (w/power factor = 1)
Rms calculations
From each measurement applied to calculation ±1 dgt.
Energy
Active and reactive power measurement accuracies ±1 dgt.
Power factor
From each measurement applied to calculation ±1 dgt.
Frequency
±0.5% rdg. (with 90 to 520 V sine wave input)
Demand value
Active and reactive power measurement accuracies ±1 dgt.
Demand quantity
Active and reactive power measurement accuracies ±1 dgt.
Conditions of
guaranteed accuracy
After 30 minute warm-up, with 50/60 Hz sine wave input
voltage to earth 400V or less
Temperature and humidity
for guaranteed accuracy
23˚C ±5˚C (73 ± 9°F), 80%RH or less
(applies to all specifications unless otherwise noted)
Display area
of guaranteed accuracy Effective measurement range
Period of guaranteed acuracy Accuracy guaranteed for 1 year
Post-adjustment accuracy guaranteed for 6 months
Real-time clock accuracy Within ±0.3 sec/day (with power on, within specified
operating temperature and humidity ranges)
Temperature characteristic Within ±0.1% f.s./ ˚C (except 23 ±5°C)
Effect of external
magnetic field
Within ±1.5% f.s.
(in a magnetic field of 400 A/m rms AC, 50/60 Hz)
Effect of radiated,
radio-frequency,
electromagnetic field
Within ±5% f.s.
for voltage and active power at 10 V/m
Range
Total display range
Effective measurement range
Total display range
Effective peak
Minimum
Minimum
Maximum
Maximum
Range
Voltage
400V Range
5.0V
90.0V
520.0V
520.0V
±750Vpeak
Current
5A Range
0.0200A
0.2500A
5.5000A
6.5000A
±20Apeak
10A Range
0.040A
0.500A
11.000A
13.000A
±40Apeak
50A Range
0.200A
2.500A
55.000A
65.000A
±200Apeak
100A Range
0.40A
5.00A
110.00A
130.00A
±400Apeak
500A Range
2.00A
25.00A
550.00A
650.00A
±1000Apeak
Measurement Range Configurations
Current Display and Effective Measurement Ranges
typical
Combined Accuracy
PW3365-20 + clamp sensors
AC FLEXIBLE CURRENT SENSOR CT9667-01, -02, -03 (500A)
Voltage Connection
Current
50.00A
100.00A
500.00A
400.0V
1P2W
20.000kW
40.000kW
200.00kW
1P3W
1P3W1U
3P3W2M
3P3W3M
40.000kW
80.000kW
400.00kW
3P4W
60.000kW
120.00kW
600.00kW
AC FLEXIBLE CURRENT SENSOR CT9667-01, -02, -03 (5kA)
Voltage Connection
Current
500.00A
1.0000kA
5.0000kA
400.0V
1P2W
200.00kW
400.00kW
2.0000MW
1P3W
1P3W1U
3P3W2M
3P3W3M
400.00kW
800.00kW
4.0000MW
3P4W
600.00kW
1.2000MW
6.0000MW