How to Use Voltage Sensors
3.2 How to Use Voltage Sensors
How to affix the voltage sensors properly
Affix each sensor to an insulated cable to be measured so that the mark
points to the cable. The sensor can be affixed to a cable with diameters
ranging from 6 mm to 30 mm.
DANGER
To prevent an electric shock, do not touch any areas
beyond the barrier while the instrument is in use.
Example: For thick
insulated cables
Barrier
Insulated cable
Insulated cable
Mark
Insulated cable
Barrier
Mark
Marks
Example: For thin insulated wires
Insulated cable
Insulated cable
Mark
Barrier
Barrier
Marks
Insulated cable
Mark