4.3 Starting and Stopping Recording
43
3
4
Chapter 4 M
easure
ment a
n
d Analy
sis
Depending on the selected recording interval, recording start is automatically delayed
until the next convenient clock time.
Recording cannot start when the battery is depleted. When the battery
becomes exhausted during recording, recording stops.
See:
"2.1 Installing (or Replacing) the Battery" ( p.17)
Automatic Recording Start at Convenient Times
Recording Interval
Recording Start Time
1 sec.
00 to 59 s (1-second interval)
2 sec.
00 to 58 s (2-seconds interval)
5 sec.
00 to 55 s (5-seconds interval)
10 sec.
00 to 50 s (10-seconds interval)
15 sec.
00 to 45 s (15-seconds interval)
20 sec.
00 to 40 s (20-seconds interval)
30 sec.
00 to 30 s (30-seconds interval)
1 min
00 min, 00 s to 59 min, 00 s (1-minute interval)
2 min
00 min, 00 s to 58 min, 00 s (2-minutes interval)
5 min
00 min, 00 s to 55 min, 00 s (5-minutes interval)
10 min
00 min, 00 s to 50 min, 00 s (10-minutes interval)
15 min
00 min, 00 s to 45 min, 00 s (15-minutes interval)
20 min
00 min, 00 s to 40 min, 00 s (20-minutes interval)
30 min
00 min, 00 s to 30 min, 00 s (30-minutes interval)
60 min
00 h, 00 min, 00 s to 23 h, 00 min, 00 s (1-hour interval)
At 12:01:00, you press the
REC
button.
The
[REC]
indicator in the display turns on,
but recording does not start yet.
At 12:10:00
Recording starts.
In the case of instantaneous value recording,
the number of data soon becomes 1.
In the case of statistical value recording, the
number of data becomes 1 at 12:20:00.
Example: When the button is pushed to start recording at 12:01:00, and
the recording interval is 10 minutes
12:01:00
Recording Start Operation
12:10:00
Actual Recording Start
If you try starting at 12:00 but are one minute late.
Actual recording starts at 12:10, which is considered the
next convenient time.
HIOKI LR5011B980-04
Содержание LR5011
Страница 2: ...HIOKI LR5011B980 04...
Страница 6: ...Contents iv HIOKI LR5011B980 04...
Страница 66: ...4 8 Printing Recorded Data 60 HIOKI LR5011B980 04...
Страница 82: ...6 4 Extracting Data 76 HIOKI LR5011B980 04...
Страница 86: ...7 2 Changing the Connection Monitoring Method and Logger Settings Displays 80 HIOKI LR5011B980 04...
Страница 102: ...9 4 Error Displays 96 HIOKI LR5011B980 04...
Страница 108: ...Index Index 4 HIOKI LR5011B980 04...
Страница 109: ...HIOKI LR5011B980 04...
Страница 110: ...HIOKI LR5011B980 04...
Страница 111: ...HIOKI LR5011B980 04...
Страница 112: ...HIOKI LR5011B980 04...