50
Input/Output/Measurement Specifications
Comparison display
Displays the difference from the measurement
voltage with the auto-hold enabled for the first
time or immediately after reset.
Valid only in the auto-hold state.
Displayed range: −100 V to 100 V
Maximum count: 100 counts
Overrange display:
[>100 V]
or
[<−100 V]
is
displayed.
Auto-hold
Setting method:
Enable by placing the instrument’s
MEASURE
key in the raised position or press and hold the
MEASURE
key on a test-lead with remote switch.
Functionality:
When both the below conditions are met,
measured value is automatically held.
• Measurement voltage: 20 V or higher
• Fluctuation range of measurement voltage for
1 second: 20 V or less
Auto-hold update: A signal meeting the above
Auto-hold conditions is input after a measurement
voltage less than 20 V is input.
(2) BPD TEST mode
Measurement
item
Bypass diode comparator judgment/Bypass route
resistor/Open-circuit voltage/Short-circuit current/
Applied current
Displayed item
Screen 1: Bypass route resistor, open-circuit voltage,
short-circuit current, and applied current
Screen 2: The number of bypass diode measurements
Measurement
object
Open-circuit voltage: 1000 V DC or less
Rated current: 2 A to 12 A DC
Crystal system string (Strings with a breakdown voltage
of 100 V DC or less cannot be measured.)
Measurement
method
Short-circuit and pulse voltage application