45
Using the
:FETCh?
Command during Continuous Measurement with Internal Triggering
This is the simplest method for exporting measured values. It is ideal when measurement (tact) time is not limited,
and when external synchronization is not needed. After connecting to the measurement target, wait for twice the
measurement time before exporting the measured value.
Using the
:READ?
Command while Continuous Measurement is Disabled
Use this method to measure (and export) synchronously with the controller (PC, PLC) or external trigger
signal. Measurement time can be minimized.
Measurement
Processing
Measured Value
Update
Command
Processing
Response
Measuring
Measuring
Measuring
Measuring
Measuring
Measurement
value
Measurement
value
Measurement
value
Measurement
value
Measurement
value
Returns the last measured value to
the PC
Measurement
Processing
Measured Value
Update
Command
Processing
Response
Measurement
Processing
Measured Value
Update
Command
Processing
Response
Measuring
Measuring
Measuring
Measuring
Measurement
value
Measurement
value
Measurement
value
Measurement
value
Internal Trigger Case
External Trigger
Idle State
Idle State
Trigger Wait
State
TRIG Signal Applied
Start Single Measurement
Returns the measured value after
single measurement to the PC
Start Single Measurement
Returns the measured value after
single measurement to the PC