11
Register Reading and Writing
Register
Read
Write
Status Byte Register
*STB?
-
Service Request Enable Register
*SRE?
*SRE
Standard Event Status Register
*ESR?
-
Standard Event Status Enable Register
*ESE?
*ESE
Event Status Register 0
:ESR0?
-
Event Status Enable Register 0
:ESE0?
:ESE0
Event Status Register 1
:ESR1?
-
Event Status Enable Register 1
:ESE1?
:ESE1
Initialization Items
Initialization Method
Item
At
Power-on
*RST
Command
:SYSTem:R
ESET
Command
*CLS
Command
Factory
Default
RS-232C setting (baud rate)
-
-
-
-
9600
Device-specific functions (range, etc.)
-
●*1
●
-
●
Output Queue
●
-
-
-
●
Input buffer
●
-
-
-
●
Status Byte Register
●
-
-
●
●
Event registers
●*2
-
-
●
●
Enable register
●
-
-
-
●
Current path
●
-
-
-
●
Headers on/off
OFF
-
-
-
OFF
*1. Except the zero adjustment value and the saved data.
*2. Except the PON bit (bit 7).
Local Function
The Remote state is entered during communication.
[RMT]
is displayed in the measurement display and operation
keys are disabled.
Canceling the Remote state
• Pressing the
[LOCAL]
key on the front panel cancels the Remote state and enables key operations.
• Sending
:SYSTem:LOCal
command can also cancel the Remote state.
LOCAL