29
Query Zero Adjustment Value (SPOT)
Syntax
Query
:ADJust:DATA:SPOT?
<V / R / RV>
Response
<Voltage adjusted value>, <Resistance adjusted value>, <Reactance
adjusted value>
Description
Query
<V> =
<R> =
<RV> =
Queries the voltage adjusted value.
Queries the resistance/reactance adjusted value at the currently
set frequency.
Queries the voltage adjusted value, and the
resistance/reactance adjusted value at the currently set
frequency.
Response
•
Response for
:ADJust:DATA:ALL? V
<Voltage adjusted value>
•
Response for
:ADJust:DATA:ALL? R
<Resistance adjusted value>, <Reactance adjusted value>
•
Response for
:ADJust:DATA:ALL? RV
<Voltage adjusted value>, <Resistance adjusted value>, <Reactance
adjusted value>
Example
ADJ:DATA:SPOT? V
+1.10000E-03
The voltage adjusted value has been set to 1.1mV.
ADJ:DATA: SPOT? R
+2.50000E-04,+2.0000E-04
The resistance/reactance adjusted value has been set to 0.25mΩ/0.20mΩ.
ADJ:DATA: SPOT? RV
+1.10000E-03,+2.50000E-04,+2.0000E-04
The voltage adjusted value has been set to 1.1mV.
The resistance/reactance adjusted value has been set to 0.25mΩ/0.20mΩ.
Query Zero Adjustment State
Syntax
Query
:ADJust:STATe?
Response
<ON / OFF>
Description
<ON> = The state of zero adjustment execution has been set to effective.
<OFF> = The state of zero adjustment execution has been set to disabled.
Example
ADJ:STAT?
OFF
The state of zero adjustment execution has been set to disabled.
Содержание BT4560
Страница 57: ...54 2 Enter the sample program into the code editor 3 Select Save All from the File menu Click...
Страница 60: ...57...
Страница 61: ...58...
Страница 62: ...59...