57
Setting Threshold Values for the Comparator
4.3 Setting Threshold Values for the
Comparator
Threshold values can be set for the comparator (resistance warning
limit, resistance failure limit, and voltage warning limit).
Example: Threshold values for a battery with the default
values* 0.4
ȍ
and 2 V.
Resistance warning limit: 0.6
ȍ
(1.5 times of the default value)
Resistance failure limit:
0.8
ȍ
(2 times of the default value)
Voltage warning limit:
1.8 V
* The default value refers to the value of resistance and voltage for
a new battery or a battery in good condition.
Selecting a comparator number
1
Press the key for at least 1
second.
7KHFRPSDUDWRUQXPEHUÀDVKHV
Press the
COMP
key again to return
to normal measurements.
2
Select the comparator
number.
(A number from 1 to 200 can be
selected.)
3
&RQ¿UPWKHVHWWLQJV
The range setup screen is displayed.
1
2
3
4
5
6
7
Appx.
Index
www.
GlobalTestSupply
.com