51
Measuring a Battery (Inspection)
6
Use the measurement values to judge whether the battery
is worn out.
Example:
FAIL
Voltage
Resistance
PASS
FAIL
WARNING
Default
value
(2.0 V)
Current
value
(1.8 V)
Present value
(1.0 m
ȍ
)
Default value
(0.5 m
ȍ
)
As shown above, this battery needs to be replaced.
To retain
measurement values
X
Refer to "3.4 Retaining the Displayed
Values" (p. 47).
To store
measurement values
X
Measurement values can be stored
by pressing the
MEMO
key while the
values are being retained.
Refer to "5.2 Storing Data in the
Memory" (p. 66).
To load the stored
data to a computer
X
Refer to "Communications Function" (p.
83).
To set threshold
values and judge
whether the battery
is worn out
X
Based on the wear judgment values,
threshold values can be set for judging
whether the battery is worn out.
Refer to "Comparator Function
(Evaluation by Threshold Values)" (p.
55).