77
________________________________________________________________________________________________________________
6.14 Particular Commands
________________________________________________________________________________________________________________
128
64
32
16
8
4
2
1
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
PON
Unused
CME
EXE
DDE
QYE
Unused Unused
Standard event status register (SESR)
6.14 Particular Commands
*
CLS
Clears the status byte register and the event registers.
*
ESR?
Queries the contents of the standard event status register (SESR).
*
IDN?
Queries manufacturer's name, model name, and software version.
Syntax
*
CLS
Function
Clears all the event registers (SESR, ESR0, ESR1) associated with the bits of the status
byte register. Accordingly, also clears the status byte register.
This has no effect upon the output queue.
Error
If the data parameters are set after this command, a command error occurs.
Syntax
*
ESR?
Function
Returns the contents of the standard event status register (SESR) as a numerical value in
NR1 format between 0 and 255, and then clears standard event status register.
No header is affixed to the response message.
Example
Response
32
Bit 5 of SESR has been set to 1.
Error
If the response message is longer than 300 bytes, a query error is generated.
Syntax
*
IDN?
Function
The response consists of the name of the manufacturer of the unit, the model name, and
the software version.
No header is affixed to the response message.
First field Manufacturer's name
Second field Model name
Third field Fixed for fifty
Fourth field Software version
Example
Response
HIOKI,3511,50,V01.00
Error
If the response message is longer than 300 bytes, a query error is generated.
Содержание 3511-50
Страница 1: ...INSTRUCTION MANUAL For 3511 50 LCR HiTESTER...
Страница 2: ......
Страница 137: ......
Страница 138: ......