5.3 General Specifications
Alarm
Criterion threshold can be set to either Hi or
Lo.
Hi : The total at every sampling is judged
within the interval.
Lo : The total is judged at every interval.
Alarm output
Alarm output × 2 CH
(Open collector: 30 VDC, 20 mAmax.)
Output is turned ON when an alarm (Hi or Lo)
occurs.
Whether to hold the alarm output can be se-
lected. The reset switch or a command can
be used to reset alarms.
Data deletion
All items of data are deleted by a command.
New data will be added to the previous data
at the start of recording.
Power outage
protection
After recovering from a power outage, the
2304-20 automatically returns to the state
held before the outage.
5.3 General Specifications
Clock accuracy
±100 ppm (Reference value at temperature
from 0 to 50°C without the communications
module)
Backup
Recorded data (saved in flash memory)
Data loss for up to 2 minutes before and after
a power outage may occur.
Communication
interface
CAN bus
Maximum rated
voltage to earth
33 Vrms, 70 VDC
Maximum input
voltage
Voltage/contact side 45 VDC
Rated supply
voltage
5 VDC±0.3 V